ST. JUDE MEDICAL, INC. ENSITE X EP SYSTEM AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number ENSITE-AMP-02 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/03/2021 |
Event Type
malfunction
|
Event Description
|
During preparation for a premature ventricular contraction (pvc) procedure with the patient in the room, the amplifier would not start, and the procedure was cancelled.The amplifier could not complete the startup cycle and an orange light kept flashing.The system was restarted several times which did not resolve the issue and the procedure was cancelled.
|
|
Manufacturer Narrative
|
One ensite x amplifier was received for evaluation.The amplifier was powered on and booted to a flashing green led status.This indicated the amplifier passed the power-on-self-test (post) and communication was established.However, the reported event was confirmed as the amplifier eventually displayed a flashing amber led status; the power on self-test was not successful.A remote connection to the amplifier¿s terminal was established, and an evaluation of the post log identified the surfacelink board at slot 6 experienced a self-test failure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the root cause was isolated to a non-functional surfacelink board.
|
|