Model Number ENSITE-AMP-02
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Event Description
During preparation for a premature ventricular contraction (pvc) procedure with the patient in the room, the amplifier would not start, and the procedure was cancelled.The amplifier could not complete the startup cycle and an orange light kept flashing.The system was restarted several times which did not resolve the issue and the procedure was cancelled.
 
Manufacturer Narrative
One ensite x amplifier was received for evaluation.The amplifier was powered on and booted to a flashing green led status.This indicated the amplifier passed the power-on-self-test (post) and communication was established.However, the reported event was confirmed as the amplifier eventually displayed a flashing amber led status; the power on self-test was not successful.A remote connection to the amplifier¿s terminal was established, and an evaluation of the post log identified the surfacelink board at slot 6 experienced a self-test failure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the root cause was isolated to a non-functional surfacelink board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE X  EP SYSTEM AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12924012
MDR Text Key281672037
Report Number2184149-2021-00394
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067034755
UDI-Public05415067034755
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K202066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENSITE-AMP-02
Device Catalogue NumberENSITE-AMP-02
Device Lot Number8068136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/03/2021
Supplement Dates Manufacturer Received03/01/2022
Supplement Dates FDA Received03/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-