• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG TREPANBOGEN HUDSON; INSTRUMENTS FOR NEUROSURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG TREPANBOGEN HUDSON; INSTRUMENTS FOR NEUROSURGERY Back to Search Results
Model Number FF056R
Medical Device Problem Codes Failure to Cycle (1142); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a trepanbogen hudson (part # ff056r) was used during a procedure performed on (b)(6) 2021.According to the complainant, during the procedure, the glove was caught in the gap of the device handle and subsequently tore.Reportedly, the glove tore twice.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Investigation results: visually, no deviation could be found at the provided device.Investigation was carried out by the q-coordinator of the production plant in charge.Excerpt of the investigation report: when checking item ff056r, the tolerance 0.15 mm (on both sides) on the crank handle was checked with a feeler gauge.If the crank handle is pressed on one side, 0.15 mm of feeler gauge goes straight into the gap that is present for the play.Conversely, this means that the specification has been met.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There is one similar complaint against the same lot number(s); similar complaint from same customer.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5) according to din en iso 14971 is still acceptable.Explanation and rationale: based on the investigation report of the responsible q-coordinator, the product is according to the specifications valid at the time of production.Conclusion and measures / preventive measures: based upon the investigation results, a clear root cause conclusion cannot be determined.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREPANBOGEN HUDSON
Common Device Name
INSTRUMENTS FOR NEUROSURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12924229
Report Number9610612-2021-00766
Device Sequence Number14632345
Product Code HXY
UDI-Device Identifier04038653162949
UDI-Public4038653162949
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFF056R
Device Catalogue NumberFF056R
Device Lot Number52602570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2021
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date12/03/2021
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-