| Model Number |
MI1200 SYNCHRONY PIN |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
During a checkup 7 weeks after the initial implantation affected channels were noticed.Reimplantation surgery was performed on the (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
During a checkup 7 weeks after the initial implantation affected channels were noticed.Reimplantation surgery was performed on the (b)(6) 2021.
|
| |
|