| Model Number |
82300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6) the platelet collection is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information.Corrected information is provided in b.3.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The run data file (rdf) was analyzed for this event.Root cause: run data file analysis did not show a conclusive root cause for the higher than expected wbc content in the platelet product reported for this collection.Alerts that could contribute to the wbc contamination of the platelet product, such as ¿centrifuge pressure high¿ or ¿rbc spillover¿, were not generated in this run data file.As the rbc detector cannot count the cells passing by, in order to generate a ¿potential wbc contamination detected¿ message, the rbc detector signals must see a significant change in the reflectance values.In this case, rbc detector reflectance signals did not indicate that wbcs were escaping the lrs chamber.Therefore, the run data file reported that the platelet product could be labeled as leukoreduced.It is also possible, though not conclusive, that this leukoreduction failure may be donor related.A potential reason for this donor related leukoreduction failure may be, but is not limited to, a high wbc count.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6) the platelet collection is not available for return because it was discarded by the customer.
|
| |
|