| Model Number |
CS 2200 |
| Medical Device Problem Code |
Unintended Application Program Shut Down (4032)
|
| Health Effect - Clinical Code |
Swelling/ Edema (4577)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Carestream dental has reviewed past complaints and noted that this is the only occurrence of this type of problem with the cs 2200.
|
| |
|
Event or Problem Description
|
|
A cs 2200 unit arm fell while in use and bumped a patient.The patient was not seriously injured and did not require any medical attention.The unit was manufactured in 2016 and installed in march 2016.The unit was installed using a passthrough installation.From the photographic evidence it appears the housing may have been filed down during installation.No maintenance was performed on the unit as recommended in the installation and service guide (sm762)_ed06) over the life of the unit.
|
| |
|
Additional Manufacturer Narrative
|
|
Carestream dental has reviewed past complaints and noted that this is the only occurrence of this type of problem with the cs 2200.
|
| |
|
Event or Problem Description
|
|
A cs 2200 unit arm fell while in use and bumped a patient.The patient was not seriously injured and did not require any medical attention.The unit was manufactured in 2016 and installed in march 2016.The unit was installed using a passthrough installation.From the photographic evidence it appears the housing may have been filed down during installation.No maintenance was performed on the unit as recommended in the installation and service guide (sm762)_ed06) over the life of the unit.
|
| |
|