| Model Number |
468-270 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Patient reported left side deflation and pain.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported left side deflation and pain.Healthcare professional later reported "bilateral exchange due to patient¿s concerns with the product.Device remains implanted.
|
| |
|
Search Alerts/Recalls
|