| Model Number |
1999/40 |
| Medical Device Problem Codes |
Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285); Capturing Problem (2891)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented remotely via (b)(6).Review of the transmission revealed that the atrial lead exhibited low impedance, high threshold and noise leading to oversensing and triggering auto mode switch episodes.Programming changes were performed.The patient was in stable condition.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's atrial lead had re-occurrences of noise.The lead was capped and replaced on (b)(6) 2025.The patient was stable.
|
| |
|