| Model Number |
TV-IL1416140-J |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924)
|
| Date of Event |
11/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the ovation ix stent graft system.Approximately four (4) years post initial procedure, a small type 3b endoleak with aneurysm enlargement was identified.Successful reintervention was performed with implant of an ovation ix iliac limb.The patient was reported to be well post-reintervention.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the type 3b endoleak of the right limb is unconfirmed.The sac growth of 11.3mm is confirmed.This is moderately consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status was discharged on the first post operative day in good condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: date received by manufacturer has been updated.H6: investigation finding: remove code 3233.H6: investigation conclusion: remove code 11.
|
| |
|
Search Alerts/Recalls
|