BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU BD INTIMA-II Y18GAX1.16IN PRN/EC SLM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383005
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2021
Event Type  malfunction  
Event Description
It was reported that bd intima-ii y18gax1.16in prn/ec slm leaked.The following information was provided by the initial reporter: "the nurse noticed fluid leakage from the indwelling needle tube during venipuncture exhausting.".
 
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Manufacturer Narrative
H6: investigation: a device history review was conducted for lot number 0231174.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
 
Event Description
It was reported that bd intima-ii y18gax1.16in prn/ec slm leaked.The following information was provided by the initial reporter: "the nurse noticed fluid leakage from the indwelling needle tube during venipuncture exhausting".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA-II Y18GAX1.16IN PRN/EC SLM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH  215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH   215126
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12929147
MDR Text Key287588787
Report Number3014704491-2021-00322
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/26/2023
Device Catalogue Number383005
Device Lot Number0231174
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-