| Model Number |
172537-LVISJ-C-PMA |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
Ruptured Aneurysm (4436)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformance associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for analysis.Images were not provided.The alleged product issue could not be confirmed.The instructions for use (ifu) lists rupture or perforation of aneurysm as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that an lvis jr.Stent was used to treat a basilar tip aneurysm after it was coiled.The distal tines of the stent did not open and a balloon was used to open them.During the second inflation attempt, the aneurysm ruptured.The stent was left implanted.
|
| |
|