PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0102
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the anterior communicating artery using penumbra smart coils (smart coils).During the procedure, two smart coils would not advance at all within their respective introducer sheath.Therefore, the two smart coils were removed.The procedure was completed using other coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2021-02745.
 
Additional Manufacturer Narrative
Evaluation of the first returned smart coil revealed, that the device was functional.During the functional, the pusher assembly was advanced out of its introducer sheath without an issue, and no further testing was performed.Evaluation of the second returned smart coil revealed, that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced, during advancement.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance, and no further issues occurred, after the pusher assembly mid-joint was out of the friction lock.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed, and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2021-02745.H3 other text: placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12931040
Report Number3005168196-2021-02744
Device Sequence Number17370110
Product Code HCG
UDI-Device Identifier00814548015613
UDI-Public00814548015613
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/16/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2026
Device Model Number400SMTHXSFT0102
Device Catalogue Number400SMTHXSFT0102
Device Lot NumberF105759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2021
Supplement Date Received by Manufacturer12/17/2021
Initial Report FDA Received Date12/04/2021
Supplement Report FDA Received Date01/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-