| Model Number |
CD3265-40Q |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2021-38432, related manufacturer reference number: 2017865-2021-38433.It was reported that the patient presented to the hospital for a follow-up.During examination, it was noted that there was infection.The implantable cardioverter defibrillator (icd), right atrial lead and left ventricular leads were explanted on (b)(6) 2021.The patient condition is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that there was infection.Interrogation of the device revealed the device was at elective replacement indicator when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The cause of infection could not be determined.
|
| |
|