• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL 5MM ROUND TIP 45 CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL 5MM ROUND TIP 45 CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSEAL545RH
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure there was a device failure.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4) batch #: v95552.The following information was requested, but unavailable: did the generator display any error messages and if so what were they? did the device pass the pre-test? had the device been working and then stopped? investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nseal545rh device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The jaws were found attached to the instrument.In addition no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL 5MM ROUND TIP 45 CM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key12932736
Report Number3005075853-2021-07397
Device Sequence Number4787280
Product Code GEI
UDI-Device Identifier10705036000051
UDI-Public10705036000051
Combination Product (Y/N)N
PMA/510(K) Number
K072493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberNSEAL545RH
Device Catalogue NumberNSEAL545RH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/24/2021
Initial Report FDA Received Date12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-