|
The rep was advised by the customer of an unspecified complaint on a blb-024115 failure that occurred on two different operations on different days.The rep advised that further information will follow.Operation 1: (b)(4).Operation 2: (b)(4).Updated as per complaint form received on 27/10/21 jil01: the rpg and rigid urs had been performed and flexor introduced into ureter.The flexible urs identified suspect tissue, so the scope was removed and the bigopsy backloaded as per ifu and handle attached as per ifu outside the patient on the scrub table.He cup would not open/close properly when tested.The device had been assembled by the consultant urologist, who detached the wire from the device and reassembled, however the same fault was still present despite assembling as per ifu (the wire was observed to be inside the window on releasing the activating button).Another bigopsy was opened and assembled as per ifu with a successful outcome and the case proceeded to completion no unintended part of the device remained inside the patient's body.No additional procedure was required due to this occurrence.No adverse effect on the patient was reported due to this occurrence.
|
|
Pma/510(k) # class i n/a.1.Device evaluation: 1 x blb-024115 of lot number c1834084 was returned to cirl for a lab evaluation.It was returned open and in its original packaging.2.Lab evaluation: the device involved in the complaint was evaluated in the laboratory on (b)(6) 2021.In summary the following results were observed in the lab evaluation.The s/s coiled wire at approximately 10cm was observed to be kinked.The forceps jaws not opening or closing properly with resistance felt.3.Image review: n/a.4.Documents review including ifu review: prior to distribution blb-024115 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for blb-024115 device of lot number c1834084 did not reveal any discrepancies that could have contributed to the issue.There is no evidence to suggest that this issue affects the entire lot # c1834084; upon review of complaints this failure mode has not occurred previously with this lot # c1834084.The instructions for use, ifu0034-8 which accompanies this device instructs the user in step 3 to ¿after the handle is attached, ensure the operation of the biopsy cups before surgical introduction.Biopsy cups default to the closed position.To open the biopsy cups pull the finger spool towards the thumb ring.To close, push the finger spool forward.Note: biopsy cups need to be in the closed position before surgical introduction.¿ according to instructions for use, ifu0034-8, the user is given the following advice: "if the package is opened or damaged when received, do not use.Visually inspect the product to ensure no damage is occurred.If abnormality is detected that would prohibit working condition, do not use".The instructions for use, ifu0034-8 which accompanies this device instructs the user in step 6 ¿ at the completion of the procedure, detach the handle according to instructions as follows: a) with the button depressed, retract the finger spool over the button.B) secure handle in this position by releasing the button (fig.2) the forceps shaft may now be removed from the handle.¿ and in step 7 ¿ remove bigopsy forceps from endoscope.¿ there is not sufficient evidence to suggest that the user did not follow the ifu.5.Root cause (possible): a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.The s/s coiled wire was found to be kinked at approximately 10cm and this resulted in the forceps jaws being unable to open and close properly with resistance felt.A possible cause of this kinked to the s/s coil wire could be attributed to damage sustained during preparation, use or handling of the device, however, this cannot be conclusively determined.6.Summary: complaint is confirmed as the failure was verified in the laboratory.According to the information reported, the patient did not experience any adverse effects or additional procedures due to this occurrence.Complaints of this nature will continue to be monitored for potential.
|