EXACTECH, INC. EQUINOXE; CAGE GLENOID S, POST AUG, LEFT Back to Search Results
Model Number 314-13-22
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by manufacturer? pending evaluation.Concomitant medical device(s): 310-01-44, 6744741 - eq humeral head short 44mm alpha, 300-51-45, 6141041 - 4.5mm short fa rep plt kit.
 
Event or Problem Description
As reported, approximately 2 months postop the initial r tsa this (b)(6) y/o female patient had early instability following a anatomic shoulder.She had an anteverted glenoid and an augment was used to correct this originally.Decision was made to convert to a reverse.Upon revision subscap was solid and all components appeared solidly fixed.Stem was retained.Reason for instability is unknown.Patient was last known to be in stable condition following the event.Facility policy doe not allow devices to be returned.
 
Additional Manufacturer Narrative
(h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the instability and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
CAGE GLENOID S, POST AUG, LEFT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key12933364
Report Number1038671-2021-00660
Device Sequence Number14669853
Product Code KWT
UDI-Device Identifier10885862196163
UDI-Public10885862196163
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/25/2026
Device Model Number314-13-22
Device Catalogue Number314-13-22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-