| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user fell while playing and hit his head.Subsequently, he could no longer hear with the device and abnormal measurements were observed.The user has been reimplanted on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, damage to the active electrode due to the reported trauma appears very likely.However to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
The user fell while playing and hit his head.Subsequently, he could no longer hear with the device and abnormal measurements were observed.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user fell while playing and hit his head.Subsequently, he could no longer hear with the device and abnormal measurements were observed.The user has been re-implanted on (b)(6), 2021.
|
| |
|
Search Alerts/Recalls
|