| Model Number |
0684-00-0480-01C |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complete event site name: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that after approximately two hours of intra-aortic balloon (iab), the console generated an unable to update timing alarm.The iab was removed.The patient was stable so another iab was not inserted.The patient was treated with medication.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The iab was placed on the cs300 pump and pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported event cannot be confirmed by the evaluation.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period nov-2019 through oct-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
| |
|