Catalog Number 2C8750 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of a clearlink system y-type blood/solution set separated.It was further reported, "the iv tubing at the end was actually cut off there was no port to attach to the iv".This issues was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to d9, h3, h4, h6 and h10.H10: the device was received for evaluation.A visual inspection was performed which observed the two-piece luer lock assembly came apart from tubing.Pressure and clear passage testing was performed and no blockages were noted.The reported condition was verified.The cause of the condition was due to a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|