• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that their device had an event code logged in its memory.The event code logged in the device's memory is indicative of a failure of the device to charge defibrillation energy.As a result, defibrillation therapy may not be available, if needed.There was no report of patient use associated with the reported event.
 
Additional Manufacturer Narrative
Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's therapy pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Stryker further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was due to a shorted capacitor, designator c62.
 
Event or Problem Description
The customer contacted stryker to report that their device had an event code logged in its memory.The event code logged in the device's memory is indicative of a failure of the device to charge defibrillation energy.As a result, defibrillation therapy may not be available, if needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12935321
Report Number0003015876-2021-02284
Device Sequence Number11959059
Product Code MKJ
UDI-Device Identifier00883873872079
UDI-Public00883873872079
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer12/23/2021
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date01/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/29/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-