SMITH & NEPHEW, INC. HEALICOIL PK 4.5MM W/ ULTRA TAPE BLUE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72203981
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during use of instruments inside the patient in a rotator cuff repair, the healicoil anchor failed on deployment and the threads peeled and fractured away from the center core upon insertion.The procedure was completed with a s+n back-up device.Non-significant delay was reported, and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found there are requirements for conformance and a certificate of material analysis is required.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.This case reports the two healicoil anchors failed on deployment, both anchors¿ threads peeled and fractured away from the center core upon insertion.Based on the limited information provided, the clinical root cause of the reported failure cannot be determined.It is unknown if the healicoil general use instruction for use was adhered to.The ifu warns: ¿it is the responsibility of the surgeon to determine the patient¿s bone condition, appropriately prepare the insertion site, and determine the suitability of the implant for the procedure.According to the report, the procedure was completed with a s+n back-up device with a non-significant delay and no other complications were reported.Therefore, no further clinical assessment is warranted at this time.Should any additional relevant information be provided, this complaint would be re-assessed.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL PK 4.5MM W/ ULTRA TAPE BLUE
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12936918
Report Number1219602-2021-02649
Device Sequence Number4884637
Product Code MBI
UDI-Device Identifier00885554032123
UDI-Public00885554032123
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113294
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2026
Device Model Number72203981
Device Catalogue Number72203981
Device Lot Number2065867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-