Device Problems
Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2021 |
Event Type
malfunction
|
Event Description
|
The hemostatic valve on the 5 f sheath was not secure and disconnected as they were injecting contrast at the end of the procedure.
|
|
Manufacturer Narrative
|
The sample is indicated as being returned.As of the date of this report, the sample has not been investigated.A follow-up report will be provided once the device has been reviewed.
|
|
Event Description
|
The hemostatic valve on the 5 f sheath was not secure and disconnected as they were injecting contrast at the end of the procedure.
|
|
Manufacturer Narrative
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.Quality engineer and complaint analyst reviewed the sample returned from the customer.Customer returned delivery catheter sheath.Returned product was visually inspected to confirm that valve cap was separated or detached from the valve, foam washer and hemo statics valve was present in the valve.Not found any indication glue properly apply on the valve cap during the assembly and the complaint was confirmed.The most probable root cause of this issue is manufacture error.During the assembly process, the operator is supposed to have the gluing tip pressed against the cap at an angle so as not to block the opening of the tip while dispensing the adhesive.According to the procedure test for hemostasis valve restriction by pushing the valve housing entirely onto the test pin fixture (the pin inserts through the cap and hemo seal).
|
|
Search Alerts/Recalls
|