TELEFLEX MEDICAL LMA SUPREME SIZE 4 Back to Search Results
Catalog Number 175040
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the doctor found cuff leakage when using on patient and then changed new one".No patient injury or harm reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Qn#(b)(4).The sample was returned and sent to the manufacturing site for investigation.The manufacturing site reports: the complaint description is related to cuff leakage during use.Based on the complaint sample returned from customer, the sample was able to be deflated and inflated normally.There was no air leak observed when the sample was inflated.The sample was also immersed into water for air leak test; there was no air leak observed from the device.From the returned complaint sample, no air leak was verified.The root cause was classified as "no problem found".".
 
Event or Problem Description
It was reported that "the doctor found cuff leakage when using on patient and then changed new one".No patient injury or harm reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LMA SUPREME SIZE 4
Common Device Name
LMA SUPREME
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
THE LARYNGEAL MASK COMPANY
6 battery road #07-02
singapore 04990 9
SN   049909
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key12939324
Report Number9681900-2021-00049
Device Sequence Number4962811
Product Code CAE
UDI-Device Identifier15060112318099
UDI-Public15060112318099
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2023
Device Catalogue Number175040
Device Lot NumberQMBCP5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer12/30/2021
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date01/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-