| Catalog Number |
175040 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient and then changed new one".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned and sent to the manufacturing site for investigation.The manufacturing site reports: the complaint description is related to cuff leakage during use.Based on the complaint sample returned from customer, the sample was able to be deflated and inflated normally.There was no air leak observed when the sample was inflated.The sample was also immersed into water for air leak test; there was no air leak observed from the device.From the returned complaint sample, no air leak was verified.The root cause was classified as "no problem found".".
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient and then changed new one".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|