| Model Number |
DKY069 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Foreign Body In Patient (2687)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device evaluation was anticipated but not yet begun.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
A patient had a total elbow replacement using the latitude ev system on the (b)(6) 2021 and was revised on the (b)(6) 2021 due to the screw preventing back out tab on the ulnar cap dky069 breaking and the screw backing out completely into the joint.The surgeon says the patient was doing extremely well with the elbow and was only when they got an xray he realised what had happened.The revision operation involved removing the broken ulnar cap, broken back out tab (in soft tissue) and screw (also in soft tissue).And replacing it with another size specific cap dky069.Nothing was noted in the initial surgery as being unusual and the surgeon has been implanting latitudes/ latitude ev's for about 15 years.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: please refer to d9/h3 product return and device evaluation.The reported event could be confirmed since x-rays were provided for the investigation and confirmed the reported event ¿broken back out tab (in soft tissue)¿.No other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
A patient had a total elbow replacement using the latitude ev system on the (b)(6) 2021 and was revised on the (h)(6) 2021 due to the screw preventing back out tab on the ulnar cap dky069 breaking and the screw backing out completely into the joint.The surgeon says the patient was doing extremely well with the elbow and was only when they got an xray he realised what had happened.The revision operation involved removing the broken ulnar cap, broken back out tab (in soft tissue) and screw (also in soft tissue).And replacing it with another size specific cap dky069.Nothing was noted in the initial surgery as being unusual and the surgeon has been implanting latitudes/ latitude ev's for about 15 years.
|
| |
|