DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0479-01C
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after the insertion of the intra-aortic balloon (iab), the customer could not remove the guidewire from the lumen.A new iab was inserted to continue therapy.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Additional Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and traces of blood on the exterior of the catheter.The extender tubing was also returned.The guide wire was not returned for evaluation.The guide wire was not returned for evaluation.The technician attempted to insert a laboratory 0.025¿ guide wire through the inner lumen and was able to successfully insert and remove the guide wire.No obstructions were felt.The reported event cannot be confirmed by the evaluation.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period dec-2019 through nov-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #:(b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 34CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12942409
Report Number2248146-2021-00824
Device Sequence Number4945992
Product Code DSP
UDI-Device Identifier10607567109329
UDI-Public10607567109329
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2023
Device Model Number0684-00-0479-01C
Device Catalogue Number0684-00-0474
Device Lot Number3000126196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2021
Supplement Date Received by Manufacturer12/17/2021
03/03/2022
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date01/04/2022
03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age62 YR
Patient SexMale
Patient Weight52 KG
-
-