| Model Number |
102620-67A |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2021,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The sensor was investigated, and though the issue couldn't be reproduced, the customer's complaint of an esr (ec#1) was however confirmed via dms' in-vivo data review.The root cause to this failure is potentially an insufficient in-vivo hydration of the sensor's hydrogel after insertion.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.As part of resolution, the rma was authorized to offer the user a sensor replacement.D9 device available for evaluation? yes, device received on 14 dec 2021.H3.Device evaluated by manufacturer?yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
|
| |
|