SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102620-67A
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
 
Additional Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Additional Manufacturer Narrative
The sensor was investigated, and though the issue couldn't be reproduced, the customer's complaint of an esr (ec#1) was however confirmed via dms' in-vivo data review.The root cause to this failure is potentially an insufficient in-vivo hydration of the sensor's hydrogel after insertion.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.As part of resolution, the rma was authorized to offer the user a sensor replacement.D9 device available for evaluation? yes, device received on 14 dec 2021.H3.Device evaluated by manufacturer?yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key12944843
Report Number3009862700-2021-00142
Device Sequence Number4912098
Product Code QCD
UDI-Device Identifier00817491022356
UDI-Public00817491022356
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/15/2022
Device Model Number102620-67A
Device Catalogue NumberFG-4201-00-301
Device Lot NumberWP08306
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2021
Supplement Date Received by Manufacturer11/06/2021
Initial Report FDA Received Date12/06/2021
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-