| Model Number |
EG-3890TLK |
| Medical Device Problem Codes |
Partial Blockage (1065); Device Fell (4014)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to pentax for further evaluation on service order (b)(4) where the user narrative was not confirmed.The inspection findings are as follows: passed dry leak test, air/ water nozzle glue missing, bending rubber glue cracking at insertion tube side, customer complaint not duplicated, distal body chipped, passed wet leak test the device was repaired where all defects found was remediated and return to the user on delivery (b)(4).If additional information becomes available, a supplemental report will be filed with the new information.
|
| |
|
Event or Problem Description
|
|
Pentax medical was made aware of an event which occurred in the pai region involving pentax video scope eg-3890tlk.In the event reported, the user states the primary channel is blocked.The timing of the event is unknown.There was no adverse event reported with this complaint.No other information provided with this complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: customer reported primary channel blocked.The customer stated unknown object in channel.However, there was no repair data that could determine the cause.We checked the returned unit and confirmed that the air/water nozzle missing.Based on the result, we concluded that it was caused due to the physical damage applied on the air/water nozzle.Correction information: g6: follow up #1, h2: type of follow up.H6: coding changed based on the investigation result.Additional information: h4: device manufacture date.
|
| |
|
Search Alerts/Recalls
|