Model Number E2516H
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Fire (2610); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Burn(s) (1757); Dizziness (2194); Electric Shock (2554)
Date of Event 11/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a procedure, the device had activation issue.The bovie was used by the doctor, who was shocked and unable to switch it back on.The handpiece did not turn on when the circulating nurse replaced the machine ports.Before using a new handpiece, they double checked to make sure the grounding pad was placed to the patient's body and plugged in.The surgeon was able to complete the surgery with a new handpiece, but the surgeon suffered a small burn on his thumb and felt light-headed and dizzy afterward.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: b5, d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that use of the device resulted in a burn and shock, and the device did not activate when keyed.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device had activation issue.The bovie was used by the doctor, who was shocked and unable to switch it back on.The handpiece did not turn on when the circulating nurse replaced the machine ports.Before they used a new handpiece, they double-checked to make sure the grounding pad was placed to the patient's body and plugged in.The surgeon was able to complete the surgery with a new handpiece.The surgical field was dry, but the surgeon suffered a small burn on her thumb and felt light-headed and dizzy afterward.They do not have the size of the burn.They stated that it was a small burn.The doctor declined treatment and evaluation at the time of the event.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key12951347
Report Number1717344-2021-01619
Device Sequence Number4911712
Product Code GEI
UDI-Device Identifier10884524000053
UDI-Public10884524000053
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K813071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE2516H
Device Catalogue NumberE2516H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2021
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date12/07/2021
Supplement Report FDA Received Date01/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-