| Model Number |
E2516H |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Fire (2610); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Dizziness (2194); Electric Shock (2554)
|
| Date of Event |
11/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the device had activation issue.The bovie was used by the doctor, who was shocked and unable to switch it back on.The handpiece did not turn on when the circulating nurse replaced the machine ports.Before using a new handpiece, they double checked to make sure the grounding pad was placed to the patient's body and plugged in.The surgeon was able to complete the surgery with a new handpiece, but the surgeon suffered a small burn on his thumb and felt light-headed and dizzy afterward.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that use of the device resulted in a burn and shock, and the device did not activate when keyed.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the device had activation issue.The bovie was used by the doctor, who was shocked and unable to switch it back on.The handpiece did not turn on when the circulating nurse replaced the machine ports.Before they used a new handpiece, they double-checked to make sure the grounding pad was placed to the patient's body and plugged in.The surgeon was able to complete the surgery with a new handpiece.The surgical field was dry, but the surgeon suffered a small burn on her thumb and felt light-headed and dizzy afterward.They do not have the size of the burn.They stated that it was a small burn.The doctor declined treatment and evaluation at the time of the event.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|