| Catalog Number |
03P76-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 24-nov-2021, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded suspected discrepant ionized calcium results on a patient.There was additional patient information available at the time of this report.Return product is not available for investigation.(b)(6).At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: #(b)(4).The investigation was completed on 16-dec-2021.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing for the lot met the acceptance criteria found in (b)(4) rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure for suppressed results and for points outside total allowable error (ea) for all sensors except ica.This related deficiency is being investigated in quality record (b)(4).
|
| |
|