• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE Back to Search Results
Catalog Number 03P76-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On 24-nov-2021, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded suspected discrepant ionized calcium results on a patient.There was additional patient information available at the time of this report.Return product is not available for investigation.(b)(6).At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident: #(b)(4).The investigation was completed on 16-dec-2021.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing for the lot met the acceptance criteria found in (b)(4) rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure for suppressed results and for points outside total allowable error (ea) for all sensors except ica.This related deficiency is being investigated in quality record (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EG7+ CARTRIDGE
Common Device Name
EG7+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key12954294
Report Number2245578-2021-00100
Device Sequence Number11816354
Product Code CHL
UDI-Device Identifier10054749000040
UDI-Public10054749000040
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K940918
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2022
Device Catalogue Number03P76-25
Device Lot NumberN21197
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/24/2021
Supplement Date Received by Manufacturer12/16/2021
Initial Report FDA Received Date12/07/2021
Supplement Report FDA Received Date12/17/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-