| Model Number |
461701E |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product. if additional information or the sample is received, the investigation will be reopened and responded to accordingly.
|
| |
|
Event or Problem Description
|
|
The customer reported that a nurse in the nicu found a defective orogastric (og) tube.Per additional information provided by the customer, the nurse pushed air into the og once it was inserted to ensure it was in the proper placement and could be heard in the stomach but was unable to hear the air.The rn removed the og tube from the patient and attempted to push water through the og at the sink area.The water squirted out of the top piece of the plastic just below where the syringe attaches.The water did not travel through the og, it only came out of the hole at the top.The nurse discarded and replaced the feeding tube with no issues.There was no patient harm and no change in the medical treatment/interventions given.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: investigation summary: the device history record (dhr) file could not be reviewed because the lot number was not available.Physical sample was not returned.Photos of the packaging was provided however, based on the images it is not possible to confirm the reported condition.If a sample is received at a later date the investigation will be updated.The exact root cause could not be determined at this time.Functional inspection is performed during the manufacturing process with acceptable results according to quality standard requirements.No action plan is deemed required.The current process is running according to product specifications, meeting quality acceptance criteria.The manufacturing site will continue monitoring the process for any adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|