• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 QUIK-COMBO; ELECTRODE, ELECTROCARDIOGRAPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 QUIK-COMBO; ELECTRODE, ELECTROCARDIOGRAPH Back to Search Results
Model Number 11996
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that their device would not recognize defibrillation pads were connected.The customer tested device and believes the defibrillation electrodes were the issue.In this state the device would not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
 
Additional Manufacturer Narrative
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.
 
Additional Manufacturer Narrative
Stryker evaluated the customer's lp 15 device - the electrodes were disposed of at the time of event and therefor not available for evaluation.The reported issue could not be duplicated and the cause could not be determined.After observing proper operation of the lp 15 device through functional and performance testing, the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted stryker to report that their device would not recognize defibrillation pads were connected.The customer tested device and believes the defibrillation electrodes were the issue.In this state the device would not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUIK-COMBO
Common Device Name
ELECTRODE, ELECTROCARDIOGRAPH
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12954929
Report Number0003015876-2021-02304
Device Sequence Number19912600
Product Code DRX
UDI-Device Identifier00883873784969
UDI-Public00883873784969
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K943301
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2022
Device Model Number11996
Device Catalogue Number11996-000017
Device Lot Number922451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2021
Supplement Date Received by Manufacturer12/20/2021
Initial Report FDA Received Date12/07/2021
Supplement Report FDA Received Date01/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LIFEPAK® 15 DEFIBRILLATOR/MONITOR, MJK, SN (B)(6).
Patient SexUnknown
-
-