| Model Number |
557P |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The patient had alleged to have black particles in the pipeline.There was no report of patient harm or injury.The device was returned to a third-party service center for investigation.Upon receipt, the device was found to have recorded 19451.1 blower hours and 16987 machine or operation hours.During the evaluation of the device at the service center, the device log files were successfully downloaded and reviewed in full.No actionable errors were identified in the log files.In addition, there was no evidence of foam particles or degradation.The customer complaint was not confirmed.The device was scrapped.The manufacturer has updated h6 in this report.
|
| |
|
Search Alerts/Recalls
|