BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7134
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left main coronary artery.A 2.00mm x 20mm emerge balloon catheter was advanced for dilation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left main coronary artery.A 2.00mm x 20mm emerge balloon catheter was advanced for dilation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.It was further reported that the target lesion was 80% stenosed and severely calcified.The balloon was inflated twice at 10-12 atmospheres for 3 seconds each.The balloon ruptured at 12 atmospheres.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded.At 47mm from the tip of the device, the guidewire lumen was buckled with a length of 4mm.The buckle contained a prolapse of the lumen and a puncture hole consistent with an interaction of a guidewire.Product analysis could not confirm reported burst.Analysis shows the guidewire lumen was buckled and punctured which is consistent with an interaction with a guidewire during procedural use.The balloon folds were also tight which indicates the balloon was not inflated.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left main coronary artery.A 2.00mm x 20mm emerge balloon catheter was advanced for dilation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.It was further reported that the target lesion was 80% stenosed and severely calcified.The balloon was inflated twice at 10-12 atmospheres for 3 seconds each.The balloon ruptured at 12 atmospheres.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12957945
Report Number2134265-2021-15476
Device Sequence Number5424509
Product Code LOX
UDI-Device Identifier08714729806301
UDI-Public08714729806301
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2023
Device Model Number7134
Device Catalogue Number7134
Device Lot Number0025528429
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/30/2021
Supplement Date Received by Manufacturer12/09/2021
01/18/2022
Initial Report FDA Received Date12/07/2021
Supplement Report FDA Received Date12/17/2021
01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight78 KG
-
-