A symptom that the remote button no.1 does not work well occurs during the pre-use inspection.The effect is good when pressed straight, but the effect from the side of the button is poor.The difference in effectiveness is clear compared to other scopes.Since it occurred before use, there is no health hazard to patients and medical staff.
|
Evaluation summary: we checked the returned unit and confirmed the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the insertion flexible tube (ift) scratched; however, it is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.Correction information: g6: follow up #1 h6: coding changed based on the investigation result additional information: h4: device manufacture date.
|