• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC 5.5 TI CORT FIX 7X40MM; ORTHOSIS, SPINAL PEDICLE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC 5.5 TI CORT FIX 7X40MM; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Model Number 186731740
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional product code: mnh, nkb, kwq, kwp, osh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that, procedure was revised due to back pain and connection degeneration.This report is for (1) 5.5 ti cort fix 7x40mm.This is report 5 of 7 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The complaint device was not received for investigation.A photo investigation was performed based on the images are received.The images were reviewed and the complaint condition is not confirmed.The rod does not appear to have any damage or manufacturing defects that would contribute to the complaint condition.A definitive assignable root cause could not be determined based on the provided information.As the device was not returned, an as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation, no product design issues or discrepancies were observed (based on the image) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device.Product code: 186731740.Lot number: arkcg7.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 27.08.2014.Qty: (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 TI CORT FIX 7X40MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
SENTIO (INNOVATIVE SURGICAL SOLUTIONS, LLC)
50461 west pontiac trail
wixom MI 48393
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12961911
Report Number1526439-2021-02520
Device Sequence Number19270231
Product Code MNI
UDI-Device Identifier10705034352008
UDI-Public(01)10705034352008
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number186731740
Device Catalogue Number186731740
Device Lot NumberARKCG7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2021
Supplement Date Received by Manufacturer12/13/2021
Initial Report FDA Received Date12/08/2021
Supplement Report FDA Received Date12/28/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SI POLYAXL SCREW 7 X 45MM; SI POLYAXL SCREW 7 X 45MM; SI POLYAXL SCREW 7 X 50MM; UNKNOWN CAGE/SPACER; UNKNOWN LOCKING/SET SCREWS; UNKNOWN RODS; UNKNOWN RODS; UNKNOWN SCREWS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-