| Model Number |
X SERIES |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device inappropriately shut down.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was verified and attributed to a loose positive and negative battery terminal.The battery contacts were re-tightened and secured to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|