BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7132
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 3.00 x 16 synergy drug-eluting stent was selected for use.However, during unpacking, it was noticed that the stent struts appeared to be lifted and deformed.The device was replaced with another stent.A 3.00mm x 12mm emerge balloon catheter was selected for use.However, during preparation upon introduction, the balloon shaft broke.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There was a separation of the hypotube 20cm from the strain relief.The separated ends of the hypotube were ovaled in shape indicating the device was kinked prior to separation.There were numerous hypotube kinks to the device.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded, and the tip of the device was damaged.Product analysis confirmed the reported break as there was separation to the device.There were unreported damages such as hypotube kinks, and tip damage.The damages found to the device are consistent with what can occur during preparation, procedural use, or an interaction with patient anatomy.
 
Event or Problem Description
It was reported that shaft break occurred.A 3.00 x 16 synergy drug-eluting stent was selected for use.However, during unpacking, it was noticed that the stent struts appeared to be lifted and deformed.The device was replaced with another stent.A 3.00mm x 12mm emerge balloon catheter was selected for use.However, during preparation upon introduction, the balloon shaft broke.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12962238
Report Number2134265-2021-15575
Device Sequence Number1854687
Product Code LOX
UDI-Device Identifier08714729806127
UDI-Public08714729806127
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2024
Device Model Number7132
Device Catalogue Number7132
Device Lot Number0027866515
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2021
Supplement Date Received by Manufacturer12/30/2021
Initial Report FDA Received Date12/08/2021
Supplement Report FDA Received Date01/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-