Model Number AQL-100CS
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The scale aql-100cs was not sent in for examination.Therefore, a malfunction that could fulfill the reporting criteria cannot be completely ruled out at present.Further attempts will be made to obtain more information about the complaint situation and to retrieve the units from the complainant.A follow-up report will be submitted.
 
Event or Problem Description
We have been informed of the following event: "territory manager, reports that they had an aquilex system that was giving a negative deficit during a case.The system started properly with a deficit of 100, then it began to decline to -295.Actual deficit was roughly 500 with a manual count.(b)(6) reports that the bladder was not emptied in the underbuttocks or primed, as there was no reason for this.Rep was present.No patient impact reported.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUILEX FLUID CONTROL SYSTEM
Common Device Name
HYSTEROSCOPY PUMP AND SCALE
Manufacturer (Section D)
W.O.M. WORLD OF MEDICINE GMBH
salzufer 8
berlin, 10587
GM  10587
Manufacturer (Section G)
W.O.M. WORLD OF MEDICINE GMBH
alte poststrasse 11
ludwigsstadt, 96337
GM   96337
Manufacturer Contact
ellen giese
salzufer 8
berlin, 10587
GM   10587
MDR Report Key12962838
Report Number3002914049-2021-00007
Device Sequence Number14554083
Product Code HIG
UDI-Device Identifier04056702002145
UDI-Public04056702002145
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K112642
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAQL-100CS
Device Catalogue NumberAQL-100CS
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-