| Model Number |
52-516-54-04 |
| Medical Device Problem Codes |
Off-Label Use (1494); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported resorbable pins used off-label to affix a titanium distractor footplate pulled loose from the footplate, necessitating removal and replacement of the pins with titanium screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot numbers: 33388735, 33405992.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.Based on the information provided the results conclude that the most likely root cause is use related.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Lot numbers 33388735.33405992.Typo and corrections: d3 - company name change.G1 - company representative name change.G1 - company establishment name updated only - typo.G1 - contact office - manufacturing site address typo.
|
| |
|