INTEGRA LIFESCIENCES MANSFIELD PROG VALVE INLINE; CHPV Back to Search Results
Model Number 82-3834
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported the physician was not able to reprogram a hakim valve.They also tried to reprogram the valve using two different programmers unsuccessfully.Valve malfunction was suspected and it was explanted and replaced.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported the physician was not able to reprogram a hakim valve.They also tried to reprogram the valve using two different programmers unsuccessfully.Valve malfunction was suspected and it was explanted and replaced.
 
Additional Manufacturer Narrative
Additional information received: the valve was implanted in (b)(6) 2021 to treat hydrocephalus in a 59-year-old female patient.The programming issue was detected by the end of (b)(6) 2021 when the patient had symptoms and the surgeon was not able to reprogram the valve.The valve was replaced with a hakim valve and the patient is doing fine.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The hakim valve was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.However, the possible root cause for the ¿programing¿ issue reported by the customer could be due to biological debris and protein build up interfering with the device.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE
Common Device Name
CHPV
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12964936
Report Number3013886523-2021-00553
Device Sequence Number19966854
Product Code JXG
UDI-Device Identifier10886704041832
UDI-Public10886704041832
Combination Product (Y/N)N
PMA/510(K) Number
K053107
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number82-3834
Device Catalogue Number823834
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2021
Supplement Date Received by Manufacturer12/22/2021
04/13/2022
Initial Report FDA Received Date12/08/2021
Supplement Report FDA Received Date01/15/2022
04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
-
-