• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-AT ARCADIS AVANTIC GEN 2; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-AT ARCADIS AVANTIC GEN 2; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number 10143408
Medical Device Problem Code Failure to Transmit Record (1521)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the (b)(6) gen 2 system.During an interventional procedure, the user reported a 3-5 second delay on the images from the c-arm.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the arcadis avantic gen 2 system.During an interventional procedure, the user reported a 3-5 second delay on the images from the c-arm.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.Assessment does not indicate a system failure or malfunction and no non-conformity was identified.The investigation was performed considering complaint description, cs reports, system history, and system log files.According to the customer, at the beginning of a procedure and after pressing the footswitch to release x-ray, it took 3 to 5 seconds until the image was displayed on the monitor.Once the image was displayed, the procedure could continue normally, without further delay.The service team performed several tests on the system (e.G.Function test of the footswitch) onsite, but could not reproduce the failure.The reported error coud not be confirmed and the system works as specified.As a preventive measure, the system generator was calibrated.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCADIS AVANTIC GEN 2
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
6104486461
MDR Report Key12967961
Report Number3004977335-2021-07937
Device Sequence Number4190324
Product Code OXO
UDI-Device Identifier04056869009001
UDI-Public04056869009001
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K051133
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10143408
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2021
Supplement Date Received by Manufacturer01/10/2022
Initial Report FDA Received Date12/09/2021
Supplement Report FDA Received Date01/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-