OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 0500318E |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Event Description
|
A clinic manager reported that a dialyzer blood leak occurred during a hemodialysis (hd) treatment.In a follow up with the clinic manager, it was clarified that the blood leak occurred immediately when the hd treatment started.The blood leak was described as being at the hanson line - drain.The leak was visually observed.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for blood.There was no damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was 250 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The treatment was completed with the same machine and new supplies.The complaint device was discarded and is not available to be returned to the manufacturer for evaluation.
|
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
|
Event Description
|
A clinic manager reported that a dialyzer blood leak occurred during a hemodialysis (hd) treatment.In a follow up with the clinic manager, it was clarified that the blood leak occurred immediately when the hd treatment started.The blood leak was described as being at the hanson line - drain.The leak was visually observed.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for blood.There was no damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was 250 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The treatment was completed with the same machine and new supplies.The complaint device was discarded and is not available to be returned to the manufacturer for evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A sample has not been provided for evaluation.A review of the production record was performed.There was no indication of product non-acceptance or deviation in the manufacturing process related to the complaint event.This includes non-conformances, rework, labeling, process controls, and any other occurrence in production that was potentially related to the complaint.The reported lot number passed pyrogen testing, was within sterilization dosage parameters and passed all release criteria.There is no recall associated with this complaint.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
|