| Catalog Number |
CAT00777 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foreign material.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: foreign material.Probable root cause: design: sterilization cycle not capable of reducing bioburden at worst-case locations device design not able to be sterilized sterilization tray design not able to accommodate sterilization of instruments process devices manufactured out of specification such that they represent an increased sterilization challenge application use past expected device life improper cleaning/disinfection before sterilization improper sterilization process, such as use of a cycle not validated for instruments, use of a different tray, or use of other instruments in tray user contaminates device.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text: 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material.
|
| |
|