Catalog Number 381023 |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that 10 bd insyte¿ autoguard¿ bc shielded iv catheters blood control technology had a damaged unit package that compromised sterility.The following information was provided by the initial reporter: " during inspection at the customer's side, 10 packages were found to be damaged/torn.".
|
|
Manufacturer Narrative
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 11/29/2021.H.6.Investigation: our quality engineer inspected the samples and photographs submitted for evaluation.Bd received 50 sealed 22g x 1.00in.Insyte autoguard bc pro units from lot number 1111856.In addition, five photographs were provided for investigation.A gross visual inspection of the returned units found that nine of the units had tears on the top web near the proximal end of the barrel.The tears had stains on the top and bottom side of the top web.The shape of the tear was in either a semi-circle shape or v-shape.The tears on some of the units were rolled back and up as if it the material was peeled back.No damage was noted to the returned dispenser box or the bottom web of the torn units.The reported issue was confirmed.Although the reported issue was confirmed, it could not be determined if the damage resulted from the manufacturing of the device or from the user environment.A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h.10.
|
|
Event Description
|
It was reported that 10 bd insyte¿ autoguard¿ bc shielded iv catheters blood control technology had a damaged unit package that compromised sterility.The following information was provided by the initial reporter: " during inspection at the customer's side, 10 packages were found to be damaged/torn.".
|
|