Model Number 08P1920
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
The customer stated that falsely elevated magnesium results were generated on the alinity c processing module.One example was provided.Sample id (b)(6) tested at (b)(6) and (b)(6) mmol/l on (b)(6) 2021.The customer's critical range is below 0.4 mmol/l.The sample was tested two additional times with results of (b)(6) and (b)(6) mmol/l.No adverse impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer stated that falsely elevated magnesium results were generated on the alinity c processing module.One example was provided.Sample id (b)(6) tested at 0.66, 0.39 and 0.66 mmol/l on november 16, 2021 using reagent lot 28011ud00.The customer's critical range is below 0.4 mmol/l.The sample was tested two additional times with results of 0.39 and 0.41 mmol/l.No adverse impact to patient management was reported.This incident is deemed reportable per criteria listed in q-22-01-015, edition 004 regarding discrepant test results for magnesium.When the initial result is within the normal range or higher and the repeat result was less than or equal to 0.53 mmol/l, the event is reportable.Additional data was provided by the customer and attached to cross-referenced ticket (b)(4) on december 15, 2021.For the analyzer documented in this ticket, an additional discrepant result (sample id (b)(6)) was documented in the attachment.The customer used a different alinity c magnesium reagent lot (27374ud00) on (b)(6) 2021 for that sample.A falsely elevated result of 1.06 mmol/l retested at 0.49 mmol/l.No adverse impact to patient management was reported.This sample is also deemed reportable as the initial result is within the normal range or higher and the repeat result was less than or equal to 0.53 mmol/l.No adverse impact to patient management was reported.
 
Additional Manufacturer Narrative
After further evaluation, the suspect medical device was updated on december 29, 2021.Mdr number 3005094123-2022-00019 has been submitted for the updated suspect medical device and all further information will be documented under that mdr number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C MAGNESIUM REAGENT KIT
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
100 abbott park road
dept. 09b9, lc cp01-3
abbott park, IL 60064-3537
2246682940
MDR Report Key12969076
Report Number3002809144-2021-00692
Device Sequence Number11866073
Product Code JGJ
UDI-Device Identifier00380740121839
UDI-Public00380740121839
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number08P1920
Device Catalogue Number08P19-20
Device Lot Number28011UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer12/29/2021
Initial Report FDA Received Date12/09/2021
Supplement Report FDA Received Date01/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6)
Patient SexUnknown
-
-