MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD SOFT MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD SOFT MESH; SURGICAL MESH Back to Search Results
Catalog Number 0117009
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, the soft mesh was noted to have ¿jagged¿ edges after the surgeon tailored the mesh, using curved mayo scissors.The sample was reported to have been implanted in the patient, as such is not available for evaluation.There has been no patient injury reported.Based on the available information, no conclusion can be made.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.To date this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in september, 2021.The instructions-for-use supplied with the mesh adequately describes the proper technique for implantation and states, "intact bard¿ soft mesh exhibits high burst and tensile strength.However, when custom tailoring, in special circumstances where excessive force is placed on the mesh, the following guidelines may be helpful: when cutting a notch in the mesh, a v-shape with a radiused point will withstand more force than a v-shape which comes to a sharp point." should additional information be provided, a supplemental mdr will be submitted.Remains implanted.
 
Event or Problem Description
As reported during an open inguinal hernia repair procedure on (b)(6) 2021, the bard soft mesh was cut to size with curved mayo scissors.As reported "the cross sections were sticking out poking into the tissue.This was felt as a safety concern by the physician that this ¿jagged¿ edge could potentially damage nerves in the area or the tissue." as reported, the size of the mesh after being cut is not known.It was reported that the mesh was used and remains implanted in the patient.As reported, there was no patient injury and the patient did not require any additional medical or surgical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD SOFT MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key12970678
Report Number1213643-2021-20459
Device Sequence Number11866084
Product Code FTL
UDI-Device Identifier00801741030864
UDI-Public(01)00801741030864
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K052155
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2026
Device Catalogue Number0117009
Device Lot NumberHUFV0737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2021
Initial Report FDA Received Date12/09/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient SexMale
Patient Weight92 KG
-
-