| Model Number |
P8.5-80B275 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
10/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed due to the patient hearing a "sound from the leg" and then experiencing pain.Additionally, it was reported that the nail was bent.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received and is pending evaluation.The root cause is unable to be determined at this time.A supplemental report will be submitted following the device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned precice nail revealed that the housing tube has bent at the middle of the nail.The work order was reviewed, and has confirmed the device passed all inspections per the acceptance tests.The iqc inspection data for the lot of the housing tube was reviewed and confirmed the parts met design specifications per the engineering drawings.Based on reported event information and review of the type of bend in the nail noted during evaluation, it is probable the nail bent due to excessive load generated from weight-bearing activities.Per the precice instructions for use: "the precice nail cannot withstand the stresses of full weight bearing or fatigue failure.Patients should utilize external support, and/or restrict activities as directed by the physician until consolidation occurs." a review of manufacturing records found no non-conformances or deviations for the lot and found all devices met all release criteria.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|
Search Alerts/Recalls
|