DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3013-53 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 11/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp.
The fse replaced the drive manifold and found the problem was the same.
The fse checked the compressor unit and found that the diaphragm of compressor was damaged which was responsible for vacuum and pressure generation.
To fix the issue, the 5000 hours maintenance kit must be replaced first.
Further diagnosis could be done after replacement of the 5000 hours maintenance kit.
Additional information is being requested with regard to the repair and status of the iabp.
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the cs100 intra-aortic balloon pump (iabp) had a low vacuum error.
It is unknown the circumstances under which the event occurred.
It is also unknown if there was a patient involvement.
However, there was no adverse event reported.
|
|
Search Alerts/Recalls
|
|
|