• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TET3030
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001); Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Wound Dehiscence (1154); Abdominal Pain (1685); Adhesion(s) (1695); Erosion (1750); Fistula (1862); Nausea (1970); Pain (1994); Scar Tissue (2060); Vomiting (2144); Hernia (2240); Distress (2329); Depression (2361); Obstruction/Occlusion (2422); Abdominal Distention (2601); Constipation (3274); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced recurrence, adhesions, small bowel obstruction, chronic pain, mesh erosion, dehiscence of mesh, and separated fascia.Post-operative patient treatment included revision surgery.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a incisional hernia.It was reported that after implant, the patient experienced recurrence, adhesions, small bowel obstruction, chronic pain, mesh erosion, dehiscence of mesh, separated fascia, ileus, abdominal pain, bloating, obstipation, constipation, depressed, nausea, vomiting, scarring, & stress.Post-operative patient treatment included revision surgery, ct scan, exploratory laparotomy, small bowel resection, gastrojejunostomy, insertion gastrostomy, hospitalization, iv fluids, pain medication, psychological assistance, reduction and repair of dehisced fascia, hernia repair with mesh, lysis of adhesions, mesh reapproximated, & nerve block.
 
Additional Manufacturer Narrative
Additional info: a4, b5, b7, h6 (patient code, imf code).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a incisional hernia.It was reported that after implant, the patient experienced fistula, recurrence, adhesions, small bowel obstruction, chronic pain, mesh erosion, dehiscence of mesh, separated fascia, ileus, abdominal pain, bloating, obstipation, constipation, depressed, nausea, vomiting, scarring, & stress.Post-operative patient treatment included revision surgery, ct scan, exploratory laparotomy, small bowel resection, gastrojejunostomy, insertion gastrostomy, hospitalization, iv fluids, pain medication, psychological assistance, reduction and repair of dehisced fascia, endoscopic removal of mesh and fistula, hernia repair with mesh, lysis of adhesions, mesh reapproximated, & nerve block.
 
Additional Manufacturer Narrative
Correction: h6 (added device code).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key12973278
Report Number9615742-2021-02473
Device Sequence Number4316927
Product Code FTL
UDI-Device Identifier10884521180185
UDI-Public10884521180185
Combination Product (Y/N)N
PMA/510(K) Number
K982532
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Model NumberTET3030
Device Catalogue NumberTET3030
Device Lot NumberSMF00215
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2021
Supplement Date Received by Manufacturer06/07/2023
03/12/2025
03/19/2025
Initial Report FDA Received Date12/09/2021
Supplement Report FDA Received Date06/19/2023
03/12/2025
03/20/2025
Date Device Manufactured06/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight94 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-