| Catalog Number |
4C8723 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that at least three (3) units of clearlink system y-type blood/solution set had a crack and leaks.The issue was identified during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h3, h4, h6 and h10.H10:the actual device was not available; however, a photograph of the sample packaging was provided for evaluation.A visual inspection was performed to the photograph, however, it was not possible to observe the reported condition through the packaging.Due to the nature of the returned sample no additional testing could be performed.Therefore, the reported problem could not be verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|