DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The getinge field service engineer (fse) who discovered the issue replaced the battery pack, li-ion x2 and verified that both ports were charging.The fse completed the pm and performed all functional and safety tests which passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.(b)(6).
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) that the cardiosave intra-aortic balloon pump (iabp) unit had one battery that will not charge.There was no patient involvement and no adverse event was reported.
|
|
Search Alerts/Recalls
|
|
|