Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Manufacturer Narrative
The getinge field service engineer (fse) who discovered the issue replaced the battery pack, li-ion x2 and verified that both ports were charging. The fse completed the pm and performed all functional and safety tests which passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service. A supplemental report will be submitted upon completion of our investigation. (b)(6). The initial reporter named is a getinge employee who has different contact details from that of the event site. (b)(6).
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) that the cardiosave intra-aortic balloon pump (iabp) unit had one battery that will not charge. There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID, TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12976672
MDR Text Key286770355
Report Number2249723-2021-02824
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/17/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-