• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Manufacturer Narrative
The getinge field service engineer (fse) who discovered the issue replaced the battery pack, li-ion x2 and verified that both ports were charging.The fse completed the pm and performed all functional and safety tests which passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.(b)(6).
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) that the cardiosave intra-aortic balloon pump (iabp) unit had one battery that will not charge.There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12976672
MDR Text Key286770355
Report Number2249723-2021-02824
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received12/10/2021
Supplement Dates Manufacturer Received03/20/2023
Supplement Dates FDA Received03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-